Medline Industries, LP: Medical Device Recall in 2026 - (Recall #: Z-1387-2026)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2026.
Data Source: FDA.
Product Description:
Medline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number: PAIN0150C
Product Classification:
Class II
Date Initiated: January 7, 2026
Date Posted: February 25, 2026
Recall Number: Z-1387-2026
Event ID: 98329
Reason for Recall:
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Status: Ongoing
Product Quantity: 448 units
Code Information:
1) PAIN0150C, UDI-DI: 10193489819571(each), 21JBV886(case), Lot Number: 21JBV886
Distribution Pattern:
Worldwide distribution - US Nationwide.
Voluntary or Mandated: