Medline Industries, LP: Medical Device Recall in 2026 - (Recall #: Z-1404-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2026.

Data Source: FDA.
Product Description:

Medline Convenience Kits: 1) BURN PACK, Model Number: DYNJ15668R; 2) BURN PACK-LF, Model Number: DYNJ42829B; 3) BURN CARE PACK-LF, Model Number: DYNJT3437

Product Classification:

Class II

Date Initiated: January 7, 2026
Date Posted: February 25, 2026
Recall Number: Z-1404-2026
Event ID: 98329
Reason for Recall:

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Status: Ongoing
Product Quantity: 291 units
Code Information:

1) DYNJ15668R, UDI-DI: 10195327108502(each), 40195327108503(case), Lot Number: 23HMB037; 2) DYNJ42829B, UDI-DI: 10193489880151(each), 40193489880152(case), Lot Number: 22IBI084; 3) DYNJT3437, UDI-DI: 10198459325076(each), 40198459325077(case), Lot Number: 25FBK633

Distribution Pattern:

Worldwide distribution - US Nationwide.

Voluntary or Mandated:

Voluntary: Firm initiated