Medline Industries, LP: Medical Device Recall in 2026 - (Recall #: Z-1404-2026)
See the recall detail below. You can also see other recalls from the same firm in 2026.
Medline Convenience Kits: 1) BURN PACK, Model Number: DYNJ15668R; 2) BURN PACK-LF, Model Number: DYNJ42829B; 3) BURN CARE PACK-LF, Model Number: DYNJT3437
Class II
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
1) DYNJ15668R, UDI-DI: 10195327108502(each), 40195327108503(case), Lot Number: 23HMB037; 2) DYNJ42829B, UDI-DI: 10193489880151(each), 40193489880152(case), Lot Number: 22IBI084; 3) DYNJT3437, UDI-DI: 10198459325076(each), 40198459325077(case), Lot Number: 25FBK633
Worldwide distribution - US Nationwide.
Voluntary: Firm initiated