Medtest Holdings, Inc.: Medical Device Recall in 2017 - (Recall #: Z-2121-2017)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2017.
Data Source: FDA.
Product Description:
Pointe Scientific Chemistry Controls Level I and II LEVl: Amber glass vial w/PP white plastic cap. LEV2: Amber glass vial w/PP black plastic cap.
Product Classification:
Class III
Date Initiated: May 15, 2014
Date Posted: May 31, 2017
Recall Number: Z-2121-2017
Event ID: 77083
Reason for Recall:
The original mean assigned to the control too high, upon reassignment with additional data the new assigned mean was outside of the originally published range.
Status: Terminated
Product Quantity: 132919 Vol vials
Code Information:
Model Number(s): C7590-50 (Chemistry Control Level I), C7591-50 (Chemistry Control Level II), and C7592-100 (Chemistry Control Level I and II).
Distribution Pattern:
Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated