Medtest Holdings, Inc.: Medical Device Recall in 2017 - (Recall #: Z-2631-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5 Levels of 1 x 2 ml ready to use calibrator or standard The Pointe Scientific CRP (HS) Wide Range Multi-Calibrator Set is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only. The Pointe Scientific CRP (HS) Wide Range Multi-Standard Set Is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only

Product Classification:

Class III

Date Initiated: July 21, 2009
Date Posted: July 5, 2017
Recall Number: Z-2631-2017
Event ID: 77451
Reason for Recall:

Individual vials of calibrator and standard were mislabeled; specifically the units on the vial labels are incorrectly listed as mg/dl when the correct unit is mg/L.

Status: Terminated
Product Quantity: 323 vials
Code Information:

Model #'s C7568-CTL and C7568-STD Lots: C7568-STD: 717303-131. Exp 6/2009. C7568-CTL: 902701-111 and 902701-131. Exp 1/2011.

Distribution Pattern:

Worldwide Distribution - US including CA, MS, NJ, PA, HI, MD, FL, OK, NV. Foreign Distribution to CHILE

Voluntary or Mandated:

Voluntary: Firm initiated