Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF): Medical Device Recall in 2018 - (Recall #: Z-0406-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
Medtronic CareLink 2090 Programmer
Product Classification:
Class II
Date Initiated: October 2, 2018
Date Posted: November 14, 2018
Recall Number: Z-0406-2019
Event ID: 81300
Reason for Recall:
Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.
Status: Ongoing
Product Quantity: 30172 units
Code Information:
All Serial Numbers
Distribution Pattern:
Worldwide
Voluntary or Mandated:
Voluntary: Firm initiated