Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF): Medical Device Recall in 2019 - (Recall #: Z-0810-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
Medtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
Class I
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality
Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01, GTIN 00643169882959, 00763000109097, All Serial Numbers (b) Model Number SPDRL1, GTIN 00643169882966, 00763000109103, All Serial Numbers
Wordlwide
Voluntary: Firm initiated