Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF): Medical Device Recall in 2021 - (Recall #: Z-1523-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
CRT-P Quad Models (CRT-Ps): a. Percepta CRT-P MRI: W1TR01, W1TR04, W4TR01, W4TR04 b. Serena CRT-P MRI: W1TR02, W1TR05, W4TR02, W4TR05 c. Solara CRT-P MRI: W1TR03, W1TR06, W4TR03, W4TR06. implantable medical devices used to treat arrhythmias related to bradycardia, and cardiac resynchronization therapy (CRT)
Product Classification:
Class II
Date Initiated: April 6, 2021
Date Posted: May 12, 2021
Recall Number: Z-1523-2021
Event ID: 87654
Reason for Recall:
There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).
Status: Ongoing
Product Quantity: 45,411 devices
Code Information:
all serial numbers
Distribution Pattern:
Worldwide
Voluntary or Mandated:
Voluntary: Firm initiated