Medtronic Inc.: Medical Device Recall in 2018 - (Recall #: Z-3201-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Contour Plus Link 2.4, Product Catalog Number: MMT-1151SK. wireless blood glucose monitoring system Contour Plus LINK 2.4 wireless blood glucose monitoring system (glucometer, test strips and control solution) is designed for self-testing by persons with diabetes and the monitoring of glucose concentrations for quantitative glucose measurement in fresh capillary full blood removed from the user's fingertips or palms. It is only for in vitro diagnostic use.

Product Classification:

Class II

Date Initiated: May 16, 2018
Date Posted: September 26, 2018
Recall Number: Z-3201-2018
Event ID: 80685
Reason for Recall:

CONTOUR PLUS LINK 2.4 meter with an incorrect unit of measure was included into meter kits and distributed. This meter contains the incorrect unit of measure mg/dL instead of mmol/L.

Status: Terminated
Product Quantity: 2
Code Information:

Lot# MD04M051P S/N# BG1245623B and S/N# BG1245624B

Distribution Pattern:

International Distribution to Slovakia only.

Voluntary or Mandated:

Voluntary: Firm initiated