Medtronic Inc.: Medical Device Recall in 2019 - (Recall #: Z-1282-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.
Class II
The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.
UDI - (01)00643169693142(10)
Distribution was only made OUS. Countries involved are: Austria, United Arab Emirate, Australia, Belgium, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Iceland, Italy, Republic of Korea, Luxembourg, Netherlands, Norway, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Turkey, and South Africa.
Voluntary: Firm initiated