Medtronic Inc.: Medical Device Recall in 2020 - (Recall #: Z-2106-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPod Touch devices.
Class II
Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge.
Model Number CSS7200 (Guardian Connect App: iOS); Catalog Numbers: MMT-7821 (Guardian Connect Transmitter - GST4C); UPN - 00763000186333. There are no assigned lot or serial numbers.
US Nationwide distribution.
Voluntary: Firm initiated