Medtronic Inc.: Medical Device Recall in 2021 - (Recall #: Z-1608-2021)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Medtronic REF: MMT-7020C2 Guardian Sensor (3) for the following packaging: GS3 5PK 9L OUS2/ GTIN: 20763000243344 GS3 5PK 9L OUS2/ GTIN: 20763000318868 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.
Product Classification:
Class II
Date Initiated: February 26, 2021
Date Posted: May 26, 2021
Recall Number: Z-1608-2021
Event ID: 87476
Reason for Recall:
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Status: Ongoing
Product Quantity: 52,196 units
Code Information:
All Lot numbers within active shelf life
Distribution Pattern:
Worldwide distribution.
Voluntary or Mandated:
Voluntary: Firm initiated