Medtronic Inc.: Medical Device Recall in 2021 - (Recall #: Z-1612-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
Medtronic REF: MMT-7020D1 Guardian Sensor (3) for the following packaging: GS3 1PK OUS1/ GTIN: 00643169704084 GS3 1PK 9L OUS1/ GTIN: 00763000243333 GS3 1PK 9L OUS1/ GTIN: 00763000318857 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.
Class II
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
All Lot numbers within active shelf life
Worldwide distribution.
Voluntary: Firm initiated