Medtronic Inc.: Medical Device Recall in 2021 - (Recall #: Z-1613-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

Medtronic REF: MMT-7020D2 Guardian Sensor (3) for the following packaging: GS3 1PK 9L OUS2/ GTIN: 00763000243357 GS3 1PK 9L OUS2/ GTIN: 00763000318871 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.

Product Classification:

Class II

Date Initiated: February 26, 2021
Date Posted: May 26, 2021
Recall Number: Z-1613-2021
Event ID: 87476
Reason for Recall:

Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Status: Ongoing
Product Quantity: 15,162 units
Code Information:

All Lot numbers within active shelf life

Distribution Pattern:

Worldwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated