Medtronic MiniMed Inc.: Medical Device Recall in 2015 - (Recall #: Z-2260-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

MiniMed NGP 640G 1.8ml (mmol/L), Model No. MMT-1511; NGP 640G 1.8ml (mg/dL), Model No. 1512; NGP 640G PLGM 3ml (mmol/L), Model No. MMT-1711; NGP 640G PLGM 3ml (mg/dL), Model No. MMT-1712.

Product Classification:

Class II

Date Initiated: June 19, 2015
Date Posted: August 5, 2015
Recall Number: Z-2260-2015
Event ID: 70897
Reason for Recall:

Medtronic MiniMed is recalling the MiniMed 620G and 640G insulin pumps because there are certain scenarios where the set Bolus screen will not timeout, which could cause confusion by showing a bolus amount that is no longer appropriate.

Status: Terminated
Product Quantity: 1936 units
Code Information:

n/a

Distribution Pattern:

Australia, Czech Republic, Denmark, Finland, France, Ireland, Netherlands, Norway, Sweden, United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated