Medtronic MiniMed: Medical Device Recall in 2020 - (Recall #: Z-1581-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
MiniMed Insulin Pump, model # MMT-508
Product Classification:
Class II
Date Initiated: June 27, 2019
Date Posted: April 1, 2020
Recall Number: Z-1581-2020
Event ID: 83433
Reason for Recall:
The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.
Status: Ongoing
Product Quantity: 64,174 units
Code Information:
All Software Versions
Distribution Pattern:
US nationwide distribution including Puerto Rico.
Voluntary or Mandated:
Voluntary: Firm initiated