Medtronic MiniMed: Medical Device Recall in 2020 - (Recall #: Z-1590-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Medtronic MiniMed Paradigm Insulin pump, model number MMT-722K

Product Classification:

Class II

Date Initiated: June 27, 2019
Date Posted: April 1, 2020
Recall Number: Z-1590-2020
Event ID: 83433
Reason for Recall:

The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.

Status: Ongoing
Product Quantity: 17
Code Information:

All Software versions

Distribution Pattern:

US nationwide distribution including Puerto Rico.

Voluntary or Mandated:

Voluntary: Firm initiated