Medtronic Navigation, Inc.: Medical Device Recall in 2014 - (Recall #: Z-1598-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
TRAJ GUIDE KIT, 9733066, BIOPSY, INT
Product Classification:
Class II
Date Initiated: April 24, 2014
Date Posted: November 26, 2014
Recall Number: Z-1598-2014
Event ID: 68145
Reason for Recall:
Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.
Status: Terminated
Product Quantity: 26,674 total
Code Information:
Lot No.: 0066233912C 0066201113B 0066212611B 0066233912C 066209012 066209012
Distribution Pattern:
Nationwide Distribution
Voluntary or Mandated:
Voluntary: Firm initiated