Medtronic Navigation, Inc.: Medical Device Recall in 2019 - (Recall #: Z-0658-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Biopsy Needle FPU Kit, Product # 9731754, UDI: 00643169341517

Product Classification:

Class II

Date Initiated: October 30, 2019
Date Posted: December 18, 2019
Recall Number: Z-0658-2020
Event ID: 84191
Reason for Recall:

The firm has become aware that due to a manufacturing issue impacting the biopsy needle depth stop, there exist the potential that some biopsy needle depth stops, when tightened, may not securely tighten to the biopsy needle. This could result in a delay in surgery.

Status: Ongoing
Product Quantity: 12 disposable biopsy needles
Code Information:

Lot Numbers: 0009665221 0009709452 0009960538

Distribution Pattern:

US: Alabama, Alaska,Arizona,Arkansas,California,Colorado,Connecticut,Delaware,District of Columbia,Florida,Georga, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine,Maryland,Massachusetts,Michigan,Minnesota,Mississippi,Missouri,Montana,Nebraska,Nevada,New Hampshire,New Jersey,New Mexico,New York,North Carolina,North Dakota,Ohio,Oklahoma,Oregon,Pennsylvania,South Carolina,South Dakota,Tennessee,Texas,Utah,Vermont,Virginia,Washington,West Virginia,Wisconsin,Wyoming, OUS: None: Australia, Austria,Bangladesh,Belgium,Bosnia And Herzegovina,Brazil,Bulgaria,Canada,Canary Islands,China,Colombia,Costa Rica,Croatia,Czech Republic,Denmark,Ecuador,Egypt,Finland,France,Germany,Greece,Hungary,Iceland,India,Indonesia,Ireland,Israel,Italy,Japan,Jordan,Kazakhstan,Kenya,Korea, Republic Of,Latvia,Lebanon,Luxembourg,Malaysia,Martinique,Mexico,Netherlands,New Zealand,Pakistan,Philippines,Poland,Portugal,Puerto Rico,Qatar,Reunion,Romania,Russian Federation,Saudi Arabia,Serbia,Slovakia,Slovenia,South Africa,Spain,Sweden,Switzerland,Thailand,Turkey,United Arab Emirates,United Kingdom

Voluntary or Mandated:

Voluntary: Firm initiated