Medtronic Navigation, Inc.: Medical Device Recall in 2020 - (Recall #: Z-2330-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
Mazor X Surgical System Positioner Type II, REF: ASM0214-02
Product Classification:
Class II
Date Initiated: December 5, 2019
Date Posted: June 17, 2020
Recall Number: Z-2330-2020
Event ID: 85387
Reason for Recall:
There is a potential for the surgical system to detach from the operational room table unexpectedly. This could result in the surgical system falling onto the patient resulting in blunt injury [hemorrhage, hematoma, bleeding, or fracture(s)]
Status: Ongoing
Product Quantity: 177 Systems
Code Information:
All Lots
Distribution Pattern:
US: ALAR,AZ,CA,CO,CT,FL,GA,IA,ID,IL,IN,KY,MA,MI,MN,MO,NC,ND,NL,NY,OH,PA,SC,TN,TX,UT,VA,WA,WI,WV, OUS: Australia, China, India, Israel, Romania, Singapore, Thailand, United Kingdom, Switzerland, Spain, Germany, Belgium,
Voluntary or Mandated:
Voluntary: Firm initiated