Medtronic Neuromodulation: Medical Device Recall in 2013 - (Recall #: Z-1580-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Medtronic SynchroMed EL Programmable pumps, Models 8626, 8626L, 8627, 8627L, (The SynchroMed EL Pumps are supplied in 10 ml or 18 ml reservoir size.) Sterilized using ethylene oxide. The pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The drug is stored in the pump reservoir. Per a programmed prescription, the drug moves from the pump reservoir, through the pump tubing, catheter port, and catheter, to the infusion site. The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the following drugs or fluids:

Product Classification:

Class I

Date Initiated: June 3, 2013
Date Posted: July 3, 2013
Recall Number: Z-1580-2013
Event ID: 65443
Reason for Recall:

Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they SynchroMed infusion pump. An electrical short circuit in a feedthrough may present as a motor stall or low battery reset/alarm and lead to a loss of or reduction in therapy which may result in the return of und

Status: Terminated
Product Quantity: 261,109 total devices estimated implanted Worldwide
Code Information:

all SynchroMed EL pumps

Distribution Pattern:

worldwide

Voluntary or Mandated:

Voluntary: Firm initiated