Medtronic Neuromodulation: Medical Device Recall in 2021 - (Recall #: Z-0293-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2021.
Data Source: FDA.
Product Description:
Medtronic Wireless Recharger Model WR9220 included in the RS5200 Recharger Kit used by the InterStim Micro (Model 97810) implantable neurostimulator.
Product Classification:
Class II
Date Initiated: November 4, 2021
Date Posted: December 1, 2021
Recall Number: Z-0293-2022
Event ID: 88953
Reason for Recall:
The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.
Status: Ongoing
Product Quantity: 10,081 devices
Code Information:
UDI 00763000186265
Distribution Pattern:
US Nationwide, including PR. There was government distribution but no military distribution was identified.
Voluntary or Mandated:
Voluntary: Firm initiated