Medtronic Neuromodulation: Medical Device Recall in 2025 - (Recall #: Z-0996-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A CLINICIAN TABLET US CT900B PROG CT900B CLINICIAN TABLET EU -UK CT900C PROG CT900C CLINICIAN TABLET UK CT900D PROG CT900D CLINICIAN TABLET GLOBAL CT900E PROG CT900E CLINICIAN TABLET GLBL SMSNG CT900F PROG CT900F CLINICIAN TABLET GLBL SMSNG

Product Classification:

Class II

Date Initiated: November 20, 2025
Date Posted: December 31, 2025
Recall Number: Z-0996-2026
Event ID: 98028
Reason for Recall:

Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the INS.

Status: Ongoing
Product Quantity: 7667 devices
Code Information:

CP App: UDI-DI 00763000273668 all software versions prior to v.1.0.4489 are affected The implanted devices are not affected.

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Malta, Mauritius, Mexico, Namibia, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Reunion, Russia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay.

Voluntary or Mandated:

Voluntary: Firm initiated