Medtronic Sofamor Danek USA Inc: Medical Device Recall in 2012 - (Recall #: Z-0036-2013)
See the recall detail below. You can also see other recalls from the same firm in 2012.
ATLANTIS(R) Anterior Cervical Plate System VARIABLE ANGLE SCREW MODULE, REF 176-503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthopedic. This instrument is a precision device which may incorporate a measuring function and has uses as described on the label. Unless labeled for single use, this instrument may be re-used.
Class II
Lines on the screw gauge may not placed on the caddy properly which could result in improper measurement of screws. An incorrect gauge could measure a screw to be approximately 1 mm shorter than its actual length.
All lots except those starting with SY
Worldwide Distribution-USA (nationwide) and the countries of India, Netherlands, Hong Kong, S. L Korea, Singapore, Canada, Australia, Japan, and Mexico.
Voluntary: Firm initiated