Product Description:
MENTOR 2" Tall Height TE With Sutures 350cc PN: 354-9312 The MENTOR CPX4 and CPX4 with Suture Tabs Breast Tissue Expanders are used for breast reconstruction following mastectomy and are intended for temporary subcutaneous or submuscular implantation and are not intended for use beyond six months.
Product Classification:
Class II
Date Initiated: August 9, 2016
Date Posted: September 14, 2016
Recall Number: Z-2747-2016
Event ID: 74895
Reason for Recall:
Tissue expander product mix found: CPX4 350cc was placed in a carton for the CPX4 450cc
Status: Terminated
Product Quantity: 65 units
Code Information:
Lot 6981537
Distribution Pattern:
Worldwide Distribution: US (nationwide) and countries of: South Korea, Argentina, Mexico, Canada, Australia and New Zealand.
Voluntary or Mandated:
Voluntary: Firm initiated