Merge Healthcare, Inc.: Medical Device Recall in 2015 - (Recall #: Z-2347-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
Merge Hemo, 9.10, 9.20.0, 9.20.1,9.20.2, 9.30, 9.40.0, 9.40.1, 9.40.2 with Massimo PHASEIN End Tidal CO2 (EtCO2) module. Merge Hemo (formerly named HeartSuite Hemodynamics) monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure.
Class II
A system freeze-up of the Merge Hemo system, that included the PHASEIN End Tidal CO2 unit, resulted in the user needing to move the patient to another cath lab.
Merge Hemo with software versions 9.10, 9.20.0, 9.20.1, 9.20.2, 9.30, 9.40.0, 9.40.1, 9.40.2 are affected only if used with any Masimo PHASEIN EtCO2 module
Nationwide Distribution including IL, SC, WI, AL, CT, CA, LA, NM, TX, OK, GA, IA, OH, FL, CO, TN, NY, NC, MA, AR, MO, HI, ND, AK, UT, WV, WA, AZ, MI, PA, ME.
Voluntary: Firm initiated