Meridian Bioscience Inc: Medical Device Recall in 2023 - (Recall #: Z-0520-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Para-Pak Zn-PVA/10% Formalin-for the routine collection, transportation, preservation, and examination of stool specimens for intestinal parasites Catalog Number: 301012
Product Classification:
Class II
Date Initiated: October 19, 2023
Date Posted: December 20, 2023
Recall Number: Z-0520-2024
Event ID: 93350
Reason for Recall:
Plastic vials contain a manufacturing defect may cause an increased risk of leakage and presents a risk of incidental exposure to biological hazards associated with direct contact with the patient specimen.
Status: Ongoing
Product Quantity: 201 Cases
Code Information:
UDI-DI: 00840733101250 Lot Numbers: 509130Q Expiration Date: 08/11/2025
Distribution Pattern:
Nationwide Foreign: Italy
Voluntary or Mandated:
Voluntary: Firm initiated