Mevion Medical Systems, Inc.: Medical Device Recall in 2020 - (Recall #: Z-2686-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation treatment to patients
Class II
The perforated screen that divides the inside of the process water tank has a very sharp edge across the top of the Heat Exchanger leading to potential injury for servicer.
Lot #s: MEVION S250-0001 MEVION S250-0002 MEVION S250-0003 MEVION S250-0004 MEVION S250-0005 MEVION S250-0006 MEVION S250-0007 MEVION S250-0008 UDI: (01)00864366000100 (11) 131209 (01)00864366000100 (11)150526 (01)00864366000100 (11)150113 (01)00864366000100 (11)150430 (01)00864366000100 (11)160127 (01)00864366000100 (11)160603 (01)00864366000124 (11) 180220 (01)00864366000124 (11) 181219
MO, NJ, OK, FL, OH, DC, Netherlands
Voluntary: Firm initiated