Micro-X Ltd.: Medical Device Recall in 2024 - (Recall #: Z-0664-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Product Classification:

Class II

Date Initiated: May 31, 2023
Date Posted: December 18, 2024
Recall Number: Z-0664-2025
Event ID: 95603
Reason for Recall:

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Status: Ongoing
Product Quantity: 5
Code Information:

UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿

Distribution Pattern:

Worldwide and US Nationwide Distribution: NJ, PR, CA

Voluntary or Mandated:

Voluntary: Firm initiated