MicroAire Surgical Instruments, LLC: Medical Device Recall in 2016 - (Recall #: Z-1328-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2016.
Data Source: FDA.
Product Description:
MicroAire PAL¿ Cannula Product Usage - Power Aspiration Device
Product Classification:
Class II
Date Initiated: February 8, 2016
Date Posted: April 13, 2016
Recall Number: Z-1328-2016
Event ID: 73373
Reason for Recall:
MicroAire has initiated a recall on product PAL-R4011XL (40cm x 4mm Del Vecchio French Rasp cannula), due to our receipt of three (3) complaints that it bent or broke during surgery. Accordingly, the PAL-R4011XL has also been put on production hold until further notice.
Status: Terminated
Product Quantity: 145 manufactured and distributed
Code Information:
Include lot numbers/serial numbers 0913228216, 0414243005, 0215266350, 0415269875, 0515271334
Distribution Pattern:
Worldwide Distribution - US Nationwide
Voluntary or Mandated:
Voluntary: Firm initiated