Microbiologics Inc: Medical Device Recall in 2022 - (Recall #: Z-0712-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299 b. KWIK-STIK 6-pack Enterococcus faecalis derived from ATCC 51299

Product Classification:

Class II

Date Initiated: October 6, 2014
Date Posted: March 9, 2022
Recall Number: Z-0712-2022
Event ID: 89451
Reason for Recall:

This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.

Status: Terminated
Product Quantity: 1,873 units
Code Information:

a. Catalog Number 0959P, UDI: 20845357023210, Lot Number (Expiration Date): 959-74-2 (10/08/2014), 959-74-3 (10/08/2014), 959-74-7 (10/08/2014), 959-75-4 (04/22/2015), 959-75-8 (04/22/2015), 959-75-9 (04/22/2015), 959-75-14 (04/22/2015), 959-76-1 (07/08/2015), 959-77-3 (10/21/2015); b. Catalog Number 0959K, UDI: 30845357023224, Lot Number (Expiration Date): 959-74-4 (10/08/2014), 959-74-6 (10/08/2014), 959-74-8 (10/08/2014), 959-75-1 (04/22/2015), 959-75-8 (04/22/2015), 959-75-12 (04/22/2015), and 959-77-1 (10/21/2015)

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Argentina, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia/Hrvatska, Ecuador, Finland, France, Germany, Guam, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Kenya, Republic of Korea, Latvia, Macedonia, Malaysia, Netherlands, Panama, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Singapore, South Africa, Switzerland, Taiwan, Tanzania, Thailand, Turkey, Uganda, United Arab Emirates, and Vietnam.

Voluntary or Mandated:

Voluntary: Firm initiated