Microbiologics Inc: Medical Device Recall in 2025 - (Recall #: Z-1163-2025)
See the recall detail below. You can also see other recalls from the same firm in 2025.
Microbiologics Multi-Strain Human Papillomavirus (HPV) Control Panel (Inactivated Pellet), REF 8234; in-vitro diagnostic test
Class II
Internal quality control procedures were not followed correctly for the testing/release of three lots of 8234/8235. This resulted in the release of products that should have failed and been discarded. The concentration of one target (HPV 16 L1) was low enough to cause QC failures, no detection, or invalid results on some instruments and assays.
UDI/DI 70845357043077, Lot Number 8234-22
Worldwide distribution - US Nationwide and the countries of Netherlands, Sweden, UK, France, Australia, New Zealand.
Voluntary: Firm initiated