MicroPort Orthopedics Inc.: Medical Device Recall in 2023 - (Recall #: Z-0276-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
EVOLUTION MP Tibial Bases, Size 7 LEFT, REF ETPKN7SL, knee tibial base
Product Classification:
Class II
Date Initiated: October 28, 2022
Date Posted: February 22, 2023
Recall Number: Z-0276-2023
Event ID: 91102
Reason for Recall:
One confirmed incident has been received that ETPKN2PL lot 1916559, size 2 Evolution MP Tibial Base, was opened during surgery and contained ETPKN7SL lot 1916715, size 7 Evolution MP Tibial Base, in the packaging.
Status: Ongoing
Product Quantity: 22 units
Code Information:
UDI/DI 0 0192629 08293 9, Lot 1916715
Distribution Pattern:
International Distribution to countries of: Germany, Italy, France, Belgium, Austria, Greece, Romania, Japan, and P.R. China.
Voluntary or Mandated:
Voluntary: Firm initiated