Mindray DS USA, Inc. dba Mindray North America: Medical Device Recall in 2025 - (Recall #: Z-0920-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Hardware configuration of the BeneVision Central Monitoring System (CMS), marketed as BeneVision Distributed Monitoring System (DMS). Model Number: 115-050935-00 and E115-050935-00. Software Version: Audio Driver version 6.0.9484.1 HP Pro Mini 400 G9 Desktop PC with Realtek HD Audio Driver Officially Released by HP (audio driver version 6.0.9484.1)

Product Classification:

Class II

Date Initiated: November 18, 2025
Date Posted: December 24, 2025
Recall Number: Z-0920-2026
Event ID: 97969
Reason for Recall:

When the Worstation of the BeneVision DMS has a specific hardware configuration, the computer may experience audio playback failure or screen freezing.

Status: Ongoing
Product Quantity: 906 units
Code Information:

Audio Driver Version: 6.0.9484.1. UDI-DI: 06936415972417. HP Pro Mini 400 G9 Desktop PC. Realtek HD Audio Driver Released by HP. 1. WS host package (mini PC). Part Number 115-050935-00. Quantity 902 2. Exch, Benevision Mini PC Workstation. Part Number E115-050935-00. Quantity 4

Distribution Pattern:

Worldwide distribution - US Nationwide and the country of Canada.

Voluntary or Mandated:

Voluntary: Firm initiated