Mizuho OSI: Medical Device Recall in 2016 - (Recall #: Z-0792-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
Mizuho OSI Trios, Table Base; Voltage: 100-240V ~ 50/60 Hz Amps: 5,0-5.0A; Model 7803 and 7803AP Orthopaedic: Operating room tablebase - an electronically powered mobile operating table is created, designed for temporary (<24 hours) support and positioning of a patient in a prone, supine, or lateral position depending on the specific table top utilized.
Class II
Table Bases may have been assembled incorrectly, missing a retaining ring from the head end assembly. Over time, the parts could go out of alignment, making clutch engagement or disengagement difficult. If the clutch cannot disengage, the user would be unable to rotate the patient from supine to prone.
Model numbers 7803, 7803AP. Serial numbers 228-261.
US, South Africa, Japan and United Arab Emirates.
Voluntary: Firm initiated