Mizuho OSI: Medical Device Recall in 2016 - (Recall #: Z-0792-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

Mizuho OSI Trios, Table Base; Voltage: 100-240V ~ 50/60 Hz Amps: 5,0-5.0A; Model 7803 and 7803AP Orthopaedic: Operating room tablebase - an electronically powered mobile operating table is created, designed for temporary (<24 hours) support and positioning of a patient in a prone, supine, or lateral position depending on the specific table top utilized.

Product Classification:

Class II

Date Initiated: December 14, 2015
Date Posted: February 17, 2016
Recall Number: Z-0792-2016
Event ID: 72937
Reason for Recall:

Table Bases may have been assembled incorrectly, missing a retaining ring from the head end assembly. Over time, the parts could go out of alignment, making clutch engagement or disengagement difficult. If the clutch cannot disengage, the user would be unable to rotate the patient from supine to prone.

Status: Terminated
Product Quantity: 33
Code Information:

Model numbers 7803, 7803AP. Serial numbers 228-261.

Distribution Pattern:

US, South Africa, Japan and United Arab Emirates.

Voluntary or Mandated:

Voluntary: Firm initiated