mo-Vis BVBA: Medical Device Recall in 2024 - (Recall #: Z-0713-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Scoot Control R-net, Product Number P015-61. The Scoot Control is intended to support a wheelchair attendant in controlling or maneuvering a power chair

Product Classification:

Class II

Date Initiated: November 28, 2023
Date Posted: January 17, 2024
Recall Number: Z-0713-2024
Event ID: 93556
Reason for Recall:

In rare cases, the lever of the throttle can get temporarily stuck during the movement and does not come back to its neutral point by itself. If this happens during usage, an unwanted movement of the chair could be the result. An unwanted movement could result in damage to the wheelchair, material damage around the chair and/or injury to the user, attendant or by-standers.

Status: Ongoing
Product Quantity: 129 units
Code Information:

UDI-DI: 05407008320119; Lot Numbers: 3304-3317 3322-3329 3440-3452 3569-3618 3830-3844 4020-4054 4170-4176 4250-4255 4321-4330

Distribution Pattern:

US States: TN, TX.

Voluntary or Mandated:

Voluntary: Firm initiated