mo-Vis BVBA: Medical Device Recall in 2025 - (Recall #: Z-2455-2025)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2025.

Data Source: FDA.
Product Description:

Multi Joystick R-net. Electrical wheelchair component.

Product Classification:

Class I

Date Initiated: July 10, 2025
Date Posted: September 10, 2025
Recall Number: Z-2455-2025
Event ID: 97354
Reason for Recall:

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

Status: Ongoing
Product Quantity: 3 units
Code Information:

Model No. P002-62; UDI-DI: 05407008320041; Serial No. 1000 to 1690.

Distribution Pattern:

US Nationwide distribution in the state of TX.

Voluntary or Mandated:

Voluntary: Firm initiated