Mobius Imaging, LLC: Medical Device Recall in 2022 - (Recall #: Z-0344-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2022.
Data Source: FDA.
Product Description:
MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1
Product Classification:
Class II
Date Initiated: November 18, 2022
Date Posted: December 7, 2022
Recall Number: Z-0344-2023
Event ID: 91178
Reason for Recall:
Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.
Status: Terminated
Product Quantity: 3
Code Information:
UDI-DI 00869346000200 Serial Numbers AIRO-0124, AIRO-0145, AIRO-0288
Distribution Pattern:
Domestic distribution to consignees in Colorado, Tennessee, and Utah.
Voluntary or Mandated:
Voluntary: Firm initiated