Motion Concepts: Medical Device Recall in 2014 - (Recall #: Z-2657-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

TRx Vertical Positioner, Classification Name: Wheelchair Component - Power Positioning System.

Product Classification:

Class II

Date Initiated: August 28, 2014
Date Posted: September 24, 2014
Recall Number: Z-2657-2014
Event ID: 69113
Reason for Recall:

Motion Concepts identified that there is a possibility of knee support failure occurring through fatigue of two attachment bolts in the Vertical Position System (TRx Vertical Positioner).

Status: Terminated
Product Quantity: US: 59 units
Code Information:

All units manufactured between June 2011 and May 2014: Serial Numbers 28583, 28597, 28604, 28607, 28629, 30334, 028380, U001050, U001083, U001544, U002020, U002326, U002386, U002403, U002411, U002489, U002502, U002505, U002509, U002510, U002519, U002716, U002922, U003577, U004011, U004018, U004073, U004079, U004145, U004214, U004400, U004404, U004472, U004485, U006042, U006234, U006235, U006260, U006295, U006296, U006353, U006378, U007126, U007138, U007163, U007239, U007245, U007279, U007288, U007407, U008556, U008631, U008698, U008840, U008896, U009548, U009569, U009579, U009707, U009718.

Distribution Pattern:

US Distribution in states of: AR, CA, CT, DE, IA, ID, IL, IN, MI, MD, MO, NE, NY, OH, PA, TN, TX, and WA.

Voluntary or Mandated:

Voluntary: Firm initiated