Murata Vios, Inc.: Medical Device Recall in 2025 - (Recall #: Z-1069-2025)
See the recall detail below. You can also see other recalls from the same firm in 2025.
muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate
Class II
During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.
Soft ware version 1.57.20, UDI: +B595BSM20500/$$7V1.57.20/16D20231129Y
Worldwide - US Nationwide distribution in the states of NY, TX, OH, MD and the country of India.
Voluntary: Firm initiated