NAImco Inc dba Richmar Inc: Medical Device Recall in 2013 - (Recall #: Z-0818-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Foam electrodes with Silver-Carbon construction: P/N 202-9271, Cat No. 400-898, 2" x 2" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9281, Cat No. 202-9281, 2" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9251, Cat No. 400-893, 3" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9291, Cat No. 400-892, 3" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9371, Cat No. 400-888, 2" x 3.5" Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9491, Cat No. 400-869, 1.25" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9511, Cat No. 400-853, 3" x 5" Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1541, Cat No. 300-154, 2" x 2.25" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1561, Cat No. 300-156, 2" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1611, Cat No. 300-161, 4" x 2.25" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010, 22818 P/N 300-1631, Cat No. 300-163, 3" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 400-8701, Cat No. 400-870, 2" x 4" Oval Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.

Product Classification:

Class II

Date Initiated: November 30, 2009
Date Posted: February 20, 2013
Recall Number: Z-0818-2013
Event ID: 63927
Reason for Recall:

Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.

Status: Terminated
Product Quantity: 56980 electrodes
Code Information:

All Serial Numbers distributed between 11/01/2009 and 01/31/2010

Distribution Pattern:

Distributed in the states of GA and TN.

Voluntary or Mandated:

Voluntary: Firm initiated