Nanosphere, Inc.: Medical Device Recall in 2014 - (Recall #: Z-2682-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

The Verigene Clostridium difficile Nucleic Acid Test (CDF) is a qualitative, multiplexed in vitro diagnostic test for the rapid detection of toxin A (tcdA), toxin B (tcdB), and tcdC gene sequences of toxigenic strains Clostridium difficile and for presumptive identification of PCR ribotype 027 strains from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI). Presumptive identification of the PCR ribotype 027 strain of C. difficile is by detection of the binary toxin (cdt) gene sequence and the single base pair deletion at nucleotide 117 in the tcdC gene. The tcdC gene encodes for a negative regulator in C. difficile toxin production. The test is performed on the Verigene System and utilizes automated specimen preparation and polymerase chain reaction (PCR) amplification, combined with a nanoparticle-based array hybridization assay to detect the toxin gene sequences associated with toxin-producing C. difficile.

Product Classification:

Class II

Date Initiated: October 10, 2013
Date Posted: October 1, 2014
Recall Number: Z-2682-2014
Event ID: 69034
Reason for Recall:

Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.

Status: Terminated
Product Quantity: 80 units
Code Information:

Product Number: 20-009-022; Lot Number and Expiration Date: Lot: 100313022B, Expiration Date: 04/01/14

Distribution Pattern:

Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain.

Voluntary or Mandated:

Voluntary: Firm initiated