Natus Neurology DBA Excel Tech., Ltd. (XLTEK): Medical Device Recall in 2018 - (Recall #: Z-2570-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
XLTEK EMU40EX EEG Headbox
Product Classification:
Class II
Date Initiated: May 15, 2018
Date Posted: August 8, 2018
Recall Number: Z-2570-2018
Event ID: 80239
Reason for Recall:
Possible degradation of an internal electronic component over time which may cause the XLTEK EMU 40EX Breakout box to fail. It may feel hot and plastic case may start melting.
Status: Terminated
Product Quantity: 462 Units Total
Code Information:
Part Numbers: 10406, 002933
Distribution Pattern:
USA Distribution: AL, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, ,RI, SC, TN, TX, UT, VA, WA, WI. World wide Distribution: Canada, China, Cyprus, Germany, Hong Kong, Italy, Mexico, Saudi Arabia, Spain, United Kingdom
Voluntary or Mandated:
Voluntary: Firm initiated