Navilyst Medical, Inc., an AngioDyamics Company: Medical Device Recall in 2016 - (Recall #: Z-1986-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
BioFlo Hybrid PICC with ENDEXO and PASV Valve Technology, Catheter Kit, UPN H965459410, Rx ONLY The Xcela PICC with PASV, Xcela Hybrid PICC with PASV, BioFlo PICC with PASV and BioFlo Hybrid PICC with PASV are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients; the sampling of blood; and for power injection of contrast media. Non-valve lumens are indicated for central venous pressure monitoring
Class II
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Batch/Lots: 4806370 Use By 2016-10-31
Worldwide Distribution US (Nationwide) Internationally to AU, GR, SE, SA, IN, CA, GB, BR, BE, NL, ES, KR, HK, AE, IT and ES.
Voluntary: Firm initiated