Navilyst Medical, Inc: Medical Device Recall in 2015 - (Recall #: Z-1151-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
NAMIC Convenience Kit, UPN H749600203741, REF/Catalog No. 60020374, STERILE, Rx ONLY --- For single use only. Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
Class II
The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
Lots 4364622 and 4376537.
Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
Voluntary: Firm initiated