Nellcor Puritan Bennett Inc. (dba Covidien LP): Medical Device Recall in 2014 - (Recall #: Z-2234-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
Vital Sync" VPMP & IM, software, V2.4. (DVD), packaged on a CD or other electronically controlled device. CDs are packaged in clam shell jeweled cases within a carton box. The Vital Sync Virtual Patient Monitoring Platform is intended to be used by healthcare professionals for the following purposes: 1) To remotely consult regarding a patients status; and 2) To remotely review near real-time patient data, waveforms and alarms in order to utilize this information to aid in clinical decisions and deliver patient care in a timely manner.
Class II
Covidien is conducting a field correction of the Vital Sync Virtual Patient Monitoring Platform and Informatics Manager due to a mismatch between the alarm priority on the medical devices and the Vital Sync system.
Model # 10007101; Lot # 14A0041IN
US distribution to states of: LA, IL, MS, CO, IN, GA, CA and TX.
Voluntary: Firm initiated