Neotract Inc: Medical Device Recall in 2018 - (Recall #: Z-0598-2019)
See the recall detail below. You can also see other recalls from the same firm in 2018.
UroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R CE 0050 Manufactured By: neotract
Class II
Upon implant deployment, the Capsular Tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no implant behind. Failure to deliver a capsular tab may result in a delay in completing a treatment or an inability to complete a treatment for the patient. Use of a recalled device may lead to increased frequency of exiting known risks associated with the product, including bleeding and tissue trauma associated with delivering the needle.
Model: AN00141 (US) and AN00142 (OUS); Catalog No.: UL400-4; Lots: P37940, P37943, P37953, P37956, P37965, P37971, P37985, P37988, P37995, P38001, P38013, P38020, P38124, P38132
US: AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV,
Voluntary: Firm initiated