NeuroLogica Corporation: Medical Device Recall in 2015 - (Recall #: Z-0263-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Product Classification:

Class II

Date Initiated: September 14, 2015
Date Posted: November 18, 2015
Recall Number: Z-0263-2016
Event ID: 72171
Reason for Recall:

When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.

Status: Terminated
Product Quantity: 7 devices
Code Information:

Serial Numbers: 50EAM9BG200004X, 50EAM9BG200001A 50EAM9BG200002P, 50EAM9BG200003D, 50EAM9BG200005N, 50EAM9BG200006E, 50EAM9BG200007F

Distribution Pattern:

Distributed to the states of NC, OH, MA, and MO.

Voluntary or Mandated:

Voluntary: Firm initiated