New Star Lasers, Inc.: Medical Device Recall in 2015 - (Recall #: Z-1918-2015)
See the recall detail below. You can also see other recalls from the same firm in 2015.
StoneLight 30 Laser System (NS3000). For use in surgical procedures using open, laparoscopic and endoscopic incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in use in medical specialties including: Urology, Urinary Lithotripsy, and General Surgery.
Class II
During internal testing a situation was discovered where when changing the settings, a new setting could not be accepted and potentially result in a discrepancy in the energy delivered (higher or lower) from the value displayed on the Graphical User Interface display.
Part Number: 0010-9400. Serial numbers: AMVM01, AMVM06, AMVM07, AMVG01, AMVP05, AMVP07, AMVP13, AMVS05, AMVK02, AMVS03, AMVS07, AMVE01, AMVS06, AMVP03, AMVS08. AMVS09, AMVS10, AMVS11, AMVS12, AMVM04, AMVM08, AMVM09, AMVM05, AMVS02, AMVK05, AMVP10, AMVS11, AMVP12, AMVP14, AMVP15, AMVP16, AMVP17, AMVP18. AMVS01, AMVK01, AMVK03, AMVK04, AMVK06, AMVK07, AMVK08, AMVK09, AMVK10, AMVM02, AMVM03, AMVM10, AMVP01, AMVP02, AMVP04, AMVP06, AMVP08, AMVP09, AMVS04.
Worldwide Distribution -- US, including the states of CA, IL, AZ, FL, OH, MA, MI, and AL; and the country of the Netherlands.
Voluntary: Firm initiated