Nextremity Solutions: Medical Device Recall in 2022 - (Recall #: Z-0739-2022)
See the recall detail below. You can also see other recalls from the same firm in 2022.
Nextra CH Cannulated Hammertoe Mini Instrument Kit-indicated for small bone reconstruction limited to inter-digital repair and fusion of the lesser toes SKU: CH-MINI-KT
Class II
During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in surgical delay, in-situ removal of the implant, use of alternative surgical approach, disassembly of the implants post-operatively, and revision
Lot Number: 1578401 UDI: 00817701026389
US Nationwide distribution in the states of AZ, IL, GA, LA, MD, NY, PA, TX, WA.
Voluntary: Firm initiated