Nextremity Solutions: Medical Device Recall in 2022 - (Recall #: Z-0911-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2022.
Data Source: FDA.
Product Description:
Nextra Hammertoe Correction System 4.5 Middle and 3.2 Proximal Saw Cut SKU: NX-4532K-SC Orthopedic surgical instrument
Product Classification:
Class II
Date Initiated: February 10, 2022
Date Posted: April 20, 2022
Recall Number: Z-0911-2022
Event ID: 89764
Reason for Recall:
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
Status: Terminated
Product Quantity: 459 units
Code Information:
Lot Number: 168117318C, 168125017D UDI: 00817701020004
Distribution Pattern:
Distribution US Nationwide, Japan, and the Netherlands
Voluntary or Mandated:
Voluntary: Firm initiated